This source measure unit meets high-capacity requirements and features remote control and easy file operations.
Source and sink operation up to 110 V / 3.2 A (four-quadrant operation) Basic accuracy: ±0.02% (DC voltage generation) Sweep output at intervals as short as 100 µs Rich sweep patterns (linear, logarithmic, arbitrary) Stores up to 65,535 points of source-measure data in internal memory Remote control and FTP via Web server function (optional) Easy file operations via USB storage function Voltage and current generation/measurement range — four-quadrant operation (source/sink) up to 110 V, 3.2 A, and 60 W; 5.5-digit output and measurement resolution; voltage range 200 mV–110 V, current range 20 µA–3.2 A; maximum output current varies depending on output voltage (±3.2 A at ≤±12 V, ±2 A at ≤±30 V, ±1 A at ≤±60 V, ±0.5 A at ≤±110 V). Generation and measurement functions — constant current source, constant voltage source, voltmeter, and ammeter, combinable into numerous operating modes (VS/IM, VS/VM, IS/VM, IS/IM, VS, IS, resistance measurement); 2-wire or 4-wire system control and measurement (local/remote sensing). Diverse operating modes — DC or pulse generation with 4 modes (continuous output, linear sweep, logarithmic sweep, programmed sweep); in programmed sweep mode, users can define an arbitrary pattern; output level can be changed at a minimum interval of 100 µs. Video: no video available on the product page. Product image: https://cdn.tmi.yokogawa.com/1/281/files/GS610_F_01_M.png GS820/GS610/GS200 High Accurate Voltage/Current Source and Measurement Unit — https://cdn.tmi.yokogawa.com/LF7656-30E.pdf Firmware update procedure — https://cdn.tmi.yokogawa.com/1/6105/files/IM765501-91E.pdf